Products
EN149 automatic mask breathing resistance tester

Scope of application】

It is used to determine the inhalation resistance and exhalation resistance of respirators and masks under specified conditions. Applicable to national labor protection equipment inspection agencies and mask manufacturers to carry out related tests and inspections of ordinary masks, dust masks, medical masks, and anti-smog masks. It can be used by manufacturers, scientific research units and third-party testing institutions to conduct KP and KN anti-particle respirators and related products are tested.

Standards compliant】

EN149-2001 7.16 etc.

technical parameter】

1. Display: 7.0-inch true color touch screen, Chinese menu operation

2. Range: 0-200 L/min, repeatability 0.5%

3. Pressure range: 0-2500Pa, accuracy 0.25%

4. Ventilation volume: 30L/min 95L/min 160L/min (can be set according to needs)

5. Head mold position: 5 positions adjustable

6. Automatic statistics and printing of the maximum value, minimum value, average value, etc.;

7. Test unit: Pa,mbar

8. Automatically test and judge test results

9. Suction volume adjustment: built-in silent air source, program feedback dynamic adjustment, stable speed, no overshoot

10. European standard head die size: face length 250mm, head width 145mm

11. Head mold material: silicone

12. Data storage: ≤1000 groups

13. Power supply: AC 220V±10V 50Hz

14. Weight: 80Kg

15. Appearance: 600×700×1100mm (L×W×H)