Products
JG468 automatic mask breathing resistance tester

Scope of application

It is used to determine the inhalation resistance and exhalation resistance of respirators and masks under specified conditions. Applicable to national labor protection equipment inspection agencies and mask manufacturers to carry out related tests and inspections of ordinary masks, dust masks, medical masks, and anti-smog masks. It can be used by manufacturers, scientific research units and third-party testing institutions to conduct KP and KN anti-particle respirators and related products are tested.

Standards compliant

GB 2626-2006, GB/T 32610-2016, YY0469, YY/T0969, etc.

technical parameter

1. Display: 7.0-inch true color touch screen, Chinese menu operation

2. Range: 0-100 L/min, accuracy 3%

3. Pressure range: 0-1500Pa, accuracy 1Pa

4. The default ventilation rate: 85±1L/min (can be set according to needs)

5. Data output: automatic statistics and printing of the maximum value, minimum value, average value, etc.;

6. Suction volume adjustment: dynamic adjustment of airflow feedback, stable speed, no overshoot

7. Data storage: 1000 groups

8. Automatically test and judge the test results, and the eligibility criteria: expiratory resistance is less than 250Pa, inspiratory resistance is less than 350Pa.

9. Power supply: AC 220V±10V 50Hz

10. Weight: 130Kg

11. Appearance: 1050*750*1350mm