YG457-I type photometer method particulate matter filtration efficiency tester (oily)

Scope of application

Medical protective respirators, daily protective respirators, disposable respirators, half-face respirators, full-face respirators, protective clothing, melt blown cloth, etc. This machine is used to test the level of oily (paraffin oil) particles removed by the filter element of the respirator under certain environmental conditions.

Applicable standards

GB2626

GB19082

GB19083

GB/T32610

YY0469

EN143

EN136

EN149

EN13274-7

JIST8151

CFR42.84

 

Specifications

1. The flow range is 0-100L/min, the accuracy is ±2%, and the resolution is 0.1L/min;

2. The flow covers multiple test standards at home and abroad;

3. The pressure detection range is 0Pa1000Pa, the accuracy is 1%, and the resolution is 0.1Pa;

4. Concentration detection range: (0.001-100) mg/L, 0.001mg/L;

5. Automatically control the concentration of aerosols;

6. Automatically control the loading of aerosol particles;

7. Configure oily aerosol generator;

8. Built-in high-precision, multi-channel-laser scattering particle (aerosol particle) sensor;

9. Measuring range of particle size: 0.3mm, 0.5mm, 0.7mm, 1.0mm, 2.5mm, 5.0mm;

10. Six-channel (0.3mm, 0.5mm, 0.7mm, 1.0mm, 2.5mm, 5.0mm) particle filtration efficiency display;

11. Concentration loading amount display;

12. The software has self-cleaning function;

13. Equipped with a variety of fixtures, suitable for testing materials such as masks, respirators, filters, protective clothing, melt blown cloth, etc.; pneumatic clamp samples;

14. Built-in anti-static device;

15. Built-in HEPA filter;

16. Built-in clean air source;

 

17. Touch screen control, simple operation;

18. With thermal printer, automatically print test data;